22.07.2014 Views

Alaris® PCA Module Guide

Alaris® PCA Module Guide

Alaris® PCA Module Guide

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Alaris® <strong>PCA</strong> <strong>Module</strong> <strong>Guide</strong><br />

Syringe Loading and Set-Up<br />

WARNING: TO PREVENT UNREGULATED FLOW, CLOSE<br />

SET TUBING CLAMP BEFORE LOADING OR UNLOADING<br />

SYRINGE.<br />

Loading:<br />

1. Open syringe barrel clamp (clear piece) until it clears syringe<br />

chamber.<br />

2. Raise drive head<br />

(gray) to fully extended<br />

position.<br />

3. Insert syringe barrel<br />

flange between barrel<br />

flange grippers<br />

(see drawing).<br />

4. Lock syringe in place<br />

by closing barrel<br />

clamp.<br />

5. Twist gripper control<br />

clockwise, lower<br />

drive head, lock<br />

plunger in place with<br />

plunger grippers.<br />

Priming:<br />

1. If priming using <strong>PCA</strong><br />

module, this feature<br />

is available when viewing Infusion Mode screen during<br />

programming of <strong>PCA</strong> module. At this screen<br />

Press OPTIONS, then press PRIME SET WITH SYRINGE.<br />

2. Once tubing set is primed, close slide clamp.<br />

Note: At the start of an infusion program, the system prompts to<br />

select and confirm syringe type and size. Ensure displayed syringe<br />

manufacturer and size correctly identifies the installed syringe.<br />

Programming <strong>PCA</strong> with <strong>PCA</strong> Pause Protocol Enabled:<br />

1. Perform steps 1-10 of Initial Set-Up in previous section and<br />

continue with following steps.<br />

2. Review Clinical Advisory “Attach an SPO2 or EtCO2 <strong>Module</strong><br />

Now”.<br />

3. Press CONFIRM.<br />

Note: If a monitoring module is not attached and started,<br />

<strong>PCA</strong> Pause Protocol WILL NOT activate.<br />

4. Choose desired Infusion Mode and follow onscreen prompts.<br />

5. Press NEXT key to verify medication parameters.<br />

6. Review Clinical Advisory “<strong>PCA</strong> Pause Limits Should be<br />

Reviewed”.<br />

7. Press CONFIRM.<br />

8. Choose desired Infusion Mode and follow onscreen prompts.<br />

Change Syringe:<br />

1. Press PAUSE and close tubing clamp.<br />

2. Use key and unlock door and remove old syringe.<br />

3. Press SILENCE.<br />

4. Attach new syringe to tubing and load new syringe.<br />

5. Set key to “Program” position and close door.<br />

6. Press CHANNEL SELECT key.<br />

7. Choose correct syringe type and size.<br />

8. Press CONFIRM.<br />

9. Press RESTORE if same drug and concentration.<br />

10. Verify drug and concentration and current settings.<br />

11. Lock door and open tubing clamp.<br />

12. Review settings and press START.<br />

Change Program/Mode:<br />

1. Press CHANNEL SELECT key.<br />

2. Press PROGRAM.<br />

3. Set key to “Program” position or enter authorization code<br />

Programming <strong>Guide</strong><br />

(if enabled).<br />

WARNING: Do not prime while attached to patient! 4. Choose desired infusion mode and follow onscreen prompts.<br />

Initial Set-Up:<br />

1. Select administration set and attach set to syringe.<br />

Beginning of Shift/Summary Review:<br />

2. If priming manually, express air from administration tubing set. 1. Press CHANNEL SELECT key and verify settings.<br />

3. Load syringe with administration set attached.<br />

2. Press START key.<br />

4. Press SYSTEM ON key and select YES or NO to “New<br />

Patient”.<br />

Patient History/End of Shift/24hr History:<br />

5. Select appropriate profile.<br />

1. Press CHANNEL SELECT key.<br />

6. Press CHANNEL SELECT key and set key to “Program” 2. Press OPTIONS.<br />

position.<br />

3. Press PATIENT HISTORY.<br />

7. Press CONFIRM time setting and choose correct syringe 4. Press ZOOM key (time interval) as appropriate and review<br />

type and size.<br />

drug totals.<br />

Note: If installed syringe is not listed, press ALL SYRINGES 5. To clear patient history press CLEAR HISTORY and press<br />

and chose matching installed syringe type and size.<br />

YES or NO.<br />

8. Choose correct medication and concentration.<br />

6. To view 24 hour totals: Press 24 h Totals.<br />

9. At “Infusion Mode” screen: To Prime, press OPTIONS key. 7. Press EXIT then press START.<br />

10. Press PRIME SET WITH SYRINGE.<br />

Operator Precautions: For proper operation of the Alaris® System (formerly<br />

Medley®System) the user must be familiar with the features, disposables,<br />

11. Press and hold PRIME key to prime tubing.<br />

administration sets, set-up and programming.<br />

Note: Do not prime while attached to patient.<br />

This guide includes selected information and suggestions and is not intended<br />

12. Press EXIT when prime is complete.<br />

to be comprehensive instructions for the set-up and operation of the Alaris®<br />

13. Choose desired Infusion Mode and follow on-screen prompts. System. For complete instructions along with Warnings and Cautions, refer to<br />

Alaris® System Directions for Use (v8).<br />

14. Close and lock door and attach administration set tubing<br />

set to patient.<br />

15. Review settings and press START.<br />

page 1 page 2<br />

<strong>PCA</strong>/Monitoring Trend Data<br />

Note: This function requires use of Alaris® monitoring<br />

module(s).<br />

1. Press CHANNEL SELECT on the monitoring module.<br />

2. Press OPTIONS.<br />

3. Press <strong>PCA</strong>/Monitoring Trend data.<br />

4. To exit: Press MAIN.<br />

5. Press MAIN SCREEN.<br />

Give a Bolus Dose<br />

1. Press CHANNEL SELECT key.<br />

2. Press Bolus Dose.<br />

3. Set key to Program position or enter authorization code (if<br />

enabled).<br />

4. Enter bolus dose amount and lock door.<br />

5. Press CONFIRM.<br />

6. Review settings and press START.<br />

Stop Bolus, Loading or <strong>PCA</strong> Dose<br />

1. Press CHANNEL SELECT key.<br />

2. Press Stop Bolus/Loading or <strong>PCA</strong>.<br />

3. Press YES or NO.<br />

Note: Programmed settings will resume.<br />

Change Dose Request Cord Setting<br />

1. Press CHANNEL SELECT key.<br />

2. Press OPTIONS.<br />

3. Press “Dose Request Set-up”.<br />

4. Choose desired Dose Cord Profile. (1=light flashes,<br />

2=light on,3=light off).<br />

5. Press CONFIRM and press START.<br />

Troubleshooting<br />

Alarm/ Meaning<br />

Alerts<br />

Check<br />

Syringe<br />

<strong>PCA</strong><br />

Pause<br />

Alarm<br />

Drive Not<br />

Engaged<br />

Max Limit<br />

Reached<br />

<strong>Module</strong><br />

Enforcement<br />

Near End<br />

(NEOI)<br />

Syringe<br />

Empty<br />

Alerts and Alarms (NOTE: During alarm state - <strong>PCA</strong> module no longer infusing).<br />

Response<br />

• Plunger grippers opened during infusion and then<br />

closed. Infusion stops on affected channel.<br />

• Syringe barrel clamp opened during infusion and<br />

then closed.<br />

• Syringe plunger not captured while in idle mode. System<br />

alarms immediately to indicate potential siphoning<br />

condition. If security door is closed and syringe plunger<br />

is not captured, the system will immediately alarm.<br />

<strong>PCA</strong> infusion has paused due to a decline in respiratory<br />

status.<br />

Drive system disengaged during operation.<br />

Indicates the maximum amount of drug delivered based<br />

on Maximum Hourly Limit field.<br />

A user message will appear on the screen when the<br />

<strong>PCA</strong> module is NOT located directly to the right of the<br />

Alaris® PC unit.<br />

Near End of Infusion and remaining VTBI will alternate<br />

on screen until syringe is empty. Alert message<br />

will scroll in channel message display on the <strong>PCA</strong><br />

module. The <strong>PCA</strong> module remains functional and will<br />

continue infusion<br />

Alarm message Syringe Empty will scroll in channel<br />

message display on the <strong>PCA</strong> module<br />

created 0808<br />

Change <strong>PCA</strong> Pause Alarm Limits<br />

1. Press CHANNEL SELECT key.<br />

2. Press OPTIONS.<br />

3. Press “<strong>PCA</strong> Pause Limits”.<br />

4. Choose desired parameter and enter value.<br />

Note: If acceptable range value is not within hospital defined<br />

range a prompt is provided.<br />

5. Follow on screen prompts.<br />

6. Press CONFIRM and press START.<br />

Access Drug Event History<br />

1. Press CHANNEL SELECT key.<br />

2. Press OPTIONS then press DRUG EVENT HISTORY.<br />

Attaching and Detaching Dose Request Cord<br />

To attach the Dose Request Cord:<br />

Insert latching connector on the cord into Dose Request Cord attachment<br />

on the <strong>PCA</strong> module. The red marking on the latching connector<br />

should be aligned with the red marking on the Dose Request Cord<br />

attachment.<br />

To detach the Dose Request Cord:<br />

Hold the body of latching connector on the Dose Request Cord and pull<br />

straight away from the <strong>PCA</strong> module, without twisting or turning.<br />

Detaching <strong>Module</strong>:<br />

Use key to unlock door. Inside locate black lever and depress. At the<br />

same time, hold the <strong>PCA</strong> module and move the bottom of the <strong>PCA</strong><br />

module sideways and away from the Alaris® PC Point-of-Care unit<br />

(“Alaris® PC unit”).<br />

Securely lock plunger grippers, press CHANNEL SELECT key, and<br />

reselect syringe. Securely lock syringe barrel clamp and press RE-<br />

START key. Check for potential siphoning. Ensure administration set<br />

clamp (roller/slide) is in closed position. Securely lock plunger grippers<br />

over syringe plunger.<br />

Assess patient status per hospital policy. Press CONFIRM once patient<br />

status and monitoring values have been addressed. Press RESTART<br />

key per hospital policy. To view trigger of <strong>PCA</strong> Pause Alarm, Press<br />

CHANNEL SELECT, Press, OPTIONS, Press DRUG EVENT HIS-<br />

TORY, Press Up/Down key to view text for monitoring value causing<br />

<strong>PCA</strong> <strong>Module</strong> to pause. Press EXIT and then START.<br />

Open and close plunger grippers. Ensure syringe is properly installed.<br />

To silence safety alarm tone, press SILENCE key. <strong>PCA</strong> module will remain<br />

silent. Alarm tone will re-sound if additional dose is requested during<br />

maximum limit reached alarm<br />

Remove the <strong>PCA</strong> module and attach directly to the right of the Alaris® PC unit.<br />

To silence safety alert tone, press SILENCE key. <strong>PCA</strong> module will remain<br />

silent until Syringe Empty alarm sounds.<br />

To silence safety alarm tone, press SILENCE key. <strong>PCA</strong> module will remain<br />

silent approximately 2 minutes and will re-sound.


<strong>PCA</strong>/EtCO2 Trend Data:<br />

Note: This function requires use of Alaris® <strong>PCA</strong> module.<br />

1. Press CHANNEL SELECT key.<br />

2. Press OPTIONS.<br />

3. Select <strong>PCA</strong>/EtCO2 Trend Data. Navigate as described<br />

above in section titled Trend Data.<br />

Troubleshooting :<br />

High Priority Meaning/ Causes<br />

Alarm<br />

No Breath<br />

Detected<br />

High EtCO2<br />

Low EtCO2<br />

High RR<br />

Low RR<br />

High FiCO2<br />

Disconnect<br />

Occluded<br />

Disposable<br />

Autozero<br />

(in progress)<br />

Clearing<br />

Disposable<br />

Alaris® EtCO2 <strong>Module</strong> <strong>Guide</strong><br />

Setting Alarm Limits:<br />

4. To exit press EtCO2 Main.<br />

1. Press CHANNEL SELECT key.<br />

5. Press MAIN SCREEN.<br />

2. Press LIMITS.<br />

3. Select limit parameter to be changed.<br />

4. Enter a numeric value using keypad or up/down arrow keys. Change Waveform Height:<br />

5. Press CONFIRM.<br />

1. Press CHANNEL SELECT key.<br />

6. Press MAIN SCREEN.<br />

2. Press OPTIONS.<br />

3. Select WAVEFORM HEIGHT.<br />

4. Select 60mmHg or 99mmHg.<br />

Trend Data:<br />

5. Press MAIN SCREEN.<br />

1. Press CHANNEL SELECT key.<br />

2. Select TREND.<br />

3. Press PAGE UP and PAGE DOWN to navigate through trend Change Waveform Time Scale:<br />

data pages. To move cursor bar press up or down arrow keys. 1. Press CHANNEL SELECT key.<br />

4. Press ZOOM to change time period.<br />

2. Press OPTIONS.<br />

5. To exit press EtCO2 Main.<br />

3. Select WAVEFORM TIME SCALE.<br />

6. Press MAIN SCREEN.<br />

4. Select 5 or 10 seconds<br />

(for lower respiratory rates select 10 seconds).<br />

5. Press MAIN SCREEN.<br />

Alarms / Messages<br />

Response<br />

• Patient is not breathing<br />

• Assess patient, check disposable.<br />

• Disposable is not properly attached to patient and/or device • Consider using different disposable type<br />

• Disposable is not detecting exhaled breath (shallow breath) • Follow hospital protocol actions.<br />

• Patient has true measurement of high EtCO2<br />

• Fever or hypermetabolic state<br />

• Disposable is not properly attached to patient<br />

• Patient has true measurement of low EtCO2<br />

• Disposable not correctly attached to patient or<br />

securely connected to module<br />

Respiratory Rate is above the specified limit<br />

Respiratory Rate is below the specified limit<br />

• Patient is inspiring exhaled CO2 or disposable not<br />

properly attached to patient<br />

• O2 mask may not be properly attached (if patient<br />

is wearing an O2 mask)<br />

• O2 flow to mask may have stopped<br />

• Drapes or covers may be over patient’s face<br />

• Purging operation failed<br />

• The disposable is occluded or needs to be reset.<br />

The module is performing an autozero calibration.<br />

During this time no data is obtained<br />

Pre-Silencing Alarm:<br />

1. Press SILENCE to pre-silence monitoring alarms for 2<br />

minutes.<br />

Note: Infusion alarms will not be silenced.<br />

• Assess patient, check disposable<br />

• Compare value to baseline<br />

• Follow hospital protocol actions.<br />

• Assess patient, check disposable,<br />

• Check O2 flow<br />

• Check mask and/or drape position<br />

• Follow hospital protocol actions<br />

• Try disconnecting disposable and then reattach.<br />

• If the device again reads<br />

DISCONNECT OCCLUDED DISPOSABLE,<br />

• Obtain and attach a new disposable.<br />

Monitoring will automatically resume when completed.<br />

No intervention is necessary.<br />

The module is trying to clear the clogged disposable. If unable to clear, the module will go into a DISCONNECT<br />

If cleared, the module will automatically resume monitoring. OCCLUDED DISPOSABLE alarm.<br />

EtCO2 Waveform Examples<br />

The following are examples of common EtCO2 waveforms. The waveform trends are examples only and do not represent all potential abnormal waveforms. Analysis<br />

of these waveform trends may provide an early indication of the noted possible causes. The associated possible responses are suggestions only and are not meant<br />

to replace current clinical practice or hospital protocols. Always consult hospital protocols. Abnormal waveforms are not always associated with alarms<br />

Normal /Abnormal Waveform Clinical Findings Possible Causes Possible Responses<br />

Operator Precautions: For proper operation of the Alaris® System (formerly Medley®System) the user must be familiar with the features,<br />

disposables, administration sets, set-up and programming. This guide includes selected information and suggestions and is not intended to be<br />

comprehensive instructions for the set-up and operation of the Alaris® System. For complete instructions along with Warnings and Cautions,<br />

refer to Alaris® System Directions for Use (v8). © 2005 Cardinal Health, Inc. or one of its subsidiaries. All rights reserved. 3POC0331<br />

page 3 page 4<br />

References: 1. Capnography in the Management of the Critically Ill Patient, EducationPAK<br />

for Critical Care and Procedural Sedation - A <strong>Guide</strong> to Capnography,<br />

CD-ROM - Needham, MA Oridion Medical, 2003.<br />

2. AACN Procedure Manual for Critial Care 4th Ed. (2001). Ed. Lynn-McHale,<br />

D.J. & Carlson K.K., American Association of Critical-Care Nurses.<br />

3. Thalan’s Critical Care Nursing Diagnosis and Management 4th Ed. (2001) Ed.<br />

Urden, L.D., Stacy, K.M. & Lough, M.E., C.V. Mosby<br />

• Normal breathing, Normal EtCO2<br />

A - B: Baseline period of no CO2, End of inhalation<br />

B - C: Exhalation begins, Begin rapid rise in CO2<br />

C - D: Sustained exhalation, Alveolar plateau<br />

D: End of expiration, end tidal CO2 (EtCO2) value<br />

D - E: Inhalation, Rapid decrease in CO2<br />

Normal Waveform<br />

(Normal Ventilation; 35-45 mmHg)<br />

Always follow hospital protocols<br />

• Treat cause of increased respiratory rate<br />

• Assess ABCs (Airway, Breathing, Circulation)<br />

• Decrease pain stimulus or encourage calm<br />

• Notify RT or MD<br />

Increase in pain level or<br />

splinting area of pain<br />

• Increase in anxiety or fear<br />

• Respiratory distress or<br />

shortness of breath<br />

Hyperventilation<br />

• Rapid breathing,<br />

Low EtCO2<br />

• Always follow hospital protocols<br />

• Access ABCs<br />

• Assess sedation level<br />

• Stimulate patient<br />

• Notify RT or MD<br />

• Over medication or<br />

increased sedation<br />

• Snoring or possible obstruction<br />

Hypoventilation<br />

• Slow breathing,<br />

High EtCO2<br />

created 0808<br />

• Always follow hospital protocols<br />

• Assess ABCs<br />

• Maintain patient airway<br />

• Encourage patient to take deep breaths<br />

• Notify RT or MD<br />

• Over medication or increased<br />

sedation<br />

• Low tidal volume<br />

Hypoventilation with Shallow Breathing<br />

Slow breathing,<br />

Low EtCO2 followed by<br />

deep breath (see pointing<br />

arrow)<br />

Always follow hospital protocols<br />

• Assess ABCs<br />

• Encourage patient to take deep breaths<br />

• Perform a head tilt or chin lift;<br />

Check position of cannula<br />

• Notify RT or MD<br />

• Poor head or neck alignment<br />

• Over medication or sedate<br />

Irregular breathing, possible<br />

audible sound or snoring,<br />

EtCO2 may be above<br />

or below baseline<br />

Partial Airway Obstruction<br />

Always follow hospital protocols<br />

• Assess ABCs<br />

• Stimulate patient<br />

• Open airway<br />

• Notify RT or MD<br />

• No Breath or Apnea<br />

• Very shallow breathing<br />

• Over medication or sedate<br />

• Displaced cannula<br />

No Breath<br />

Sudden loss of EtCO2<br />

reading, Very shallow or no<br />

respiratory rate pattern<br />

observed

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!